Feature and specification of the isolation gown
| Feature | Specification |
|---|---|
| Product Definition | Non-sterile, single-use garment. Classified as a Medical Device Class I. Primary purpose is patient protection (to prevent transfer of infectious agents from wearer to patient), with secondary protection for the wearer. |
| Key Material | Non-woven materials such as SMS, SMMS, polyethylene-coated polypropylene, or equivalents. |
| Core Performance | Water resistant with a minimum Hydrostatic Head (HH) rate of 200 mm, corresponding to ANSI/AAMI PB70 Level 2. It offers limited liquid barrier protection and is not specifically tested as an infectious barrier. |
| Key Design | Full-back design with back closure. Long sleeves. Must provide coverage from neck to below the knees. Universal size with a front length of 130–145 cm and a minimum waist circumference of 130 cm. |
| Closure Options | Neck: Velcro (preferred), tape tab, or tie. Waist: Tie at the back or side. Cuffs: Cotton/poly blend knit, elastic, or thumb loop (least preferred). |
The gown is intended for two main isolation scenarios
Protective Isolation: For immune-depressed patients (e.g., burn patients, malnourished children).
Infectious Isolation: For diseases transmitted by airborne contact (droplets).